MicroSYNTHplus Specifications
PARALLEL SYNTHESIS |
| ROTOR |
PRO-6 |
PRO - 16/24 |
High-Pressure |
Large-Volume |
| Number of vessels |
6 |
16-24 |
10 |
6 |
| Vessels volume (ml) |
12-50 |
75 |
100 |
270 |
| Working volume (ml) |
2-30 |
10-35 |
10-60 |
50-160 |
| Maximum temperature (°C) |
180 |
200 |
240 |
200 |
| Maximum pressure (bar) |
15 |
30 |
100 |
10 |
| Vessel material |
Glass or Quartz |
TFM |
SCALE-UP PRESSURE REACTORS |
| Rotor volume (ml) |
500 |
1000 |
| Working volume (ml) |
50-150 |
100-500 |
| Maximum temperature (°C) |
200 |
| Maximum pressure (bar) |
14 |
8 |
| Vessel material |
TFM |
Microwave Hardware
- Dual magnetron system with pyramid-shaped diffuser for homogeneous microwave distribution in the cavity.
- Magnetron protection from reflected microwave power.
- Installed power 1.600 watt.
- Output power 1.000 watt, controlled via microprocessor.
- User-selectable continuous and pulsed microwave emission.
- Patented self-sealing pressure responsive door.
- Large microwave cavity 35 x 35 x 35 cm.
- Cavity illumination.
- Microwave cavity entirely made of 18/8 stainless steel housing with an innovative multi-layer PTFE plasma coating
applied at over 350° C.
- All hardware protected against acids/organic solvents with polymer coating both on inside and outside surfaces.
- Total of 7 micro-switches of which 4 safety interlocks to prevent microwave emission with door open.
- Exhaust located above the cavity, separate from electronics to prevent corrosion.
- Can accomodate large volume BatchSYNTH reactor.
Control Terminal

- RAM 256 MB up to 512 MB on the CPU board
- Real-time clock battery buffered
Interfaces
- COM1: RS-232, or-485
- COM2: RS-232
- External VGA Monitor connector (max. 1280x1024 resolution)
- PS/2-keyboard/mouse connector (shared)
- Ethernet
- Display Colour-TFT, 15”, 1024x768 resolution
- Touch-screen, mouse compatible
- Internal Compact Flash Card (Standard 512MB)
- External USB-Memory stick
- Power supply 100-240 V, 50-60 Hz, 1.35A
- Output: 42 watt max
|