DMA-80 analytical performance The DMA-80 produces accurate results over a wide dynamic range, on a variety of different matrices. The results on the right indicate analyses on certified reference materials with mercury concentrations ranging from 5 μg/kg to over 25 mg/kg.
DMA-80 recovery on certified reference samples.
The DMA-80 provides reproducible results over a broad range of matrices, including polymers and biological samples, and over a wide concentration range.
MEMORY EFFECT
The DMA-80 can accommodate samples sizes of up to 1.5 grams. This is particularly important for the analysis of heterogeneous materials and for samples with a very low mercury content. The graphic illustrates the results obtained analyzing NIST 2709 (San Joaquin Soil), a certified reference material with known mercury content of 1.32- 1.48 mg/kg. All results obtained on the DMA-80 are within the certified range. Analysis of NIST 2709 Soil.
The DMA-80 produces quality results regardless of the sample matrix. The instrument can be calibrated using aqueous standards and then be used in the analysis of organic and inorganic solid samples. In most cases, one method can be used to analyze a variety of different matrices making the DMA-80 extremely easy to use.
DIFFICULT SAMPLES Due to their high reactivity, the analysis of samples such as naphtha, lubricant oil and crude oil can be problematic. Typical drying/decomposition |